A dielectric test applies a voltage and watches the current. The pass criterion is "no breakdown", which is a qualitative event. The trip current is what the instrument does about it: above a set threshold, it cuts the voltage and reports a failure.
Set it too low and capacitive inrush trips the instrument on good samples. Set it too high and a real breakdown draws enough energy to damage the sample and, on a bad day, the operator.
Choosing the setting
| Sample | Typical trip setting | Reason |
| Small appliance, basic insulation | 5 mA | Leakage is a fraction of a milliamp; 5 mA catches a breakdown early |
| Appliance with EMI filter | 10 mA to 20 mA | Y capacitors draw current at 50 Hz |
| Motor with large winding capacitance | 20 mA to 50 mA | Capacitive current scales with test frequency and capacitance |
| Cable or long harness | 50 mA to 100 mA | The cable itself is a capacitor |
| Production line, repetitive | Lower setting plus ramp rate limit | Repeatability matters more than margin |
| Type test, one-off | Higher setting with current monitoring | You want to see the trend before the trip |
The rule of thumb: measure the leakage current on a known-good sample first, then set the trip at three to five times that value. If the result is below 5 mA, use 5 mA and accept the extra trips.
Why capacitive current surprises people
At 50 Hz a 10 nF capacitor draws about 0.3 mA at 1 kV. A filter with two 4.7 nF Y capacitors draws about the same. Multiply by the number of filters and the current adds up quickly, and none of it is a fault.
- DC testing removes the capacitive current. That is the main practical reason to choose DC for long cables.
- Frequency matters. Current scales linearly with frequency at the same voltage.
- Ramp slowly. A fast ramp produces an inrush that a steady-state reading would not.
- Watch the trend, not the number. A sample whose leakage climbs over ten seconds is breaking down even if it never trips.
The operator safety side
A hipot tester is a current-limited source, and the limit is what keeps a contact harmless. Instruments are built to keep the output below about 5 mA into a short for this reason; a setting of 100 mA exists for the sample, not for the person.
That is why the return lead goes to the earthed chassis and the probe is held with one hand, and why the discharge step after the test is not optional. Our dielectric testers log the leakage current continuously so the trend is visible, and the technical notes cover the settings for filtered products.
Standards referenced: IEC 60335-1 clause 13.3 and 16.3, IEC 61010-1 clause 6.8, IEC 60243-1 for the dielectric strength method.
The first hour of a production line
Set the trip current from a measured baseline: run 5 known-good units, record the leakage current at the test voltage, and set the trip at 3 times the highest value, with a floor of 5 mA. On a line running 200 units per shift, a setting that is 2 mA too low produces 6 false trips a day and teaches the operator to press reset without looking.
- Record the baseline on the station, not in a file.
- Re-baseline whenever the product or the filter changes.
- Never raise the setting to clear a trip; find the unit that caused it.