Test probe B and test probe 11 look like the same tool. Both are 12 mm in diameter, both carry a 50 mm stop face, both are used to check that hazardous live parts are not accessible. The difference is a single joint, and that joint decides which products each probe can test.
The joint is the whole point
Probe B has a jointed finger: a knuckle roughly 30 mm from the tip that articulates through 90° in one plane, with a tolerance of 0 to +10°. Probe 11 has no joint. It is a straight 12 mm rod.
The joint exists because a human finger bends. When someone reaches into a recess, the finger follows the contour — it goes in straight, then bends around a corner to reach a terminal. A straight rod cannot do that; it stops at the corner. Testing a product with probe 11 where probe B is required will report "not accessible" for a part that a real finger reaches easily.
Conversely, probe 11 is the right tool where the question is narrower: can a straight finger reach this opening? That question comes up on flat faces with small apertures, where a joint would only get in the way.
| Feature | Probe B | Probe 11 |
|---|---|---|
| Rod diameter | 12 mm | 12 mm |
| Stop face | 50 mm | 50 mm |
| Joint | 90°, 0 to +10° | none |
| Tests | recesses, corners, behind barriers | straight access through an opening |
Why the standards use both
IEC 61032 defines both probes because the two questions are genuinely different. A product standard that says "verify accessibility with test probe B" is asking whether a finger can reach the part, following the contour. A standard that says "test probe 11" is asking a flatter question, usually about an opening in a plane surface.
Where a product standard names one and not the other, the other is not a substitute. Substituting probe 11 for probe B is the more dangerous direction, because it reports "not accessible" too readily.
The force question
The applied force depends on the standard family, not the probe. IP2X checking under IEC 60529 uses 10 N. Household appliance accessibility under IEC 60335-1 uses 30 N, and plugs and socket outlets under IEC 60884-1 also use 30 N. There is no 20 N step in the base standards.
The force is applied along the probe axis, with the stop face in contact with the enclosure. On probe B the joint must be allowed to follow the contour — if the operator holds the joint straight, the test degenerates into a probe 11 test.
Checking the two probes
Probe 11 is easy to check: diameter, stop face flatness, tip radius. Probe B adds the joint, which is where most of the uncertainty lives.
Droop test. Hold the probe vertically by the handle. The tip should not sag under its own weight, or only slightly. A sagging joint bends before it reaches the part.
Rotation test. Move the joint through its full 90° and feel for binding. A stiff joint stops the probe at the corner — the same failure as using probe 11.
Angle check. The joint's permitted range is 90° with a 0 to +10° tolerance. A joint that swings past 100° is out of specification even if it feels fine.
None of these need a lab. They need thirty seconds at the bench and a written record, which is exactly the part that gets skipped.