Insulation resistance is not a material constant. It depends on the voltage applied, the time the voltage is applied, the temperature, and the humidity history of the sample. Choosing a test voltage is therefore choosing which of those effects you will see.
What the standard sets
Medical equipment at 250 V, household appliances at 500 V, and some industrial equipment at 1,000 V. The choice follows the working voltage, not the convenience of the instrument.
| Application | Test voltage | Typical minimum | Note |
| Class II appliance, basic insulation | 500 V DC | 2 MΩ | IEC 60335-1 clause 16 |
| Medical, applied part | 250 V DC | 2 MΩ | IEC 60601-1 clause 8.8 |
| Industrial control | 500 V DC | 1 MΩ | IEC 61010-1 clause 6.9 |
| High voltage equipment | 1,000 V DC | 10 MΩ | Product standard decides |
Why the reading falls with voltage
Above a certain field strength, conduction through a moist or contaminated surface becomes non-linear. Doubling the voltage can halve the resistance. That is not a measurement error; it is the difference between an ohmic path and an ionic one.
For a dry, clean surface the reading is nearly independent of voltage, which is why a large drop between two voltages is itself diagnostic: the insulation has a moisture or contamination problem that a single low-voltage reading would miss.
Time matters as much as voltage
Insulation appears to charge. The reading 15 s after the voltage is applied is lower than the reading after 60 s, and the ratio of the two is the polarisation index. A PI below 2.0 on a winding suggests moisture or contamination; above 2.0 is normal for a healthy dry winding.
Record the voltage, the electrification time and the temperature. Three numbers make the result reproducible; one number makes it a guess.
Standards referenced: IEC 60335-1 clause 16, IEC 60601-1 clause 8.8, IEC 61010-1 clause 6.9, IEC 60270 for partial discharge measurements.
Guard circuits and surface leakage
A low insulation reading is often surface leakage rather than bulk conduction. A guard terminal, connected to a conductive ring around the insulator, diverts surface current away from the measurement and shows whether the problem is on the surface or through the material.
If the reading rises by a factor of ten when the guard is connected, the problem is contamination on the surface. If it barely changes, the insulation itself has absorbed moisture.
Temperature correction
Insulation resistance roughly halves for every 10 K rise in temperature, for many common materials. A reading taken at 35 °C is therefore about a quarter of the reading at 15 °C, and comparing the two without correction is meaningless.
Either measure at a controlled temperature or apply the correction and state it. A report that quotes 50 MΩ without a temperature cannot be compared with one that quotes 200 MΩ.
Where the reading should be taken
Between live parts and accessible metal, and between live parts and the enclosure surface where the enclosure is insulating. For Class II equipment, the second measurement is the one that matters, because there is no earth to fall back on.