A hipot tester applies a high voltage and watches the current. On a resistive fault the current rises and the tester trips, which is the intended behaviour. On a long cable the same thing happens with no fault at all, because a cable with two conductors is a capacitor and a capacitor draws current while its voltage is changing.
The current a capacitor draws
I = C × dV/dt. A 100 m run of two-core cable has roughly 10 nF of capacitance. Ramping to 1,500 V in 1 s gives dV/dt of 1,500 V/s, so the charging current is 15 µA, which is negligible.
Ramp the same cable to 1,500 V in 0.1 s and the charging current becomes 150 µA, still small. The problem appears with large capacitance or with a tester that applies a step: a 10 µF load stepped to 1,500 V in 10 ms draws 1.5 A.
| Load | Capacitance | Ramp 1,500 V in 1 s | Ramp in 50 ms |
| 100 m two-core cable | 10 nF | 15 µA | 300 µA |
| 1 km multicore cable | 100 nF | 150 µA | 3 mA |
| Motor winding | 1 to 5 µF | 1.5 to 7.5 mA | 30 to 150 mA |
| Filter with Y-capacitors | 0.5 µF | 0.75 mA | 15 mA |
| Large capacitor bank | 10 µF | 15 mA | 1.5 A |
Why 500 VA is not the number that matters
The VA rating describes how much power the tester can deliver continuously. The trip is set in milliamps and responds in milliseconds. A 500 VA tester can still trip on a 30 mA setting if the load charges fast enough, because the VA rating does not limit the instantaneous current the way the trip circuit sees it.
Three fixes, in order of preference
- Slow the ramp. Most testers allow 1 to 10 s; a 5 s ramp removes nearly all capacitive trips.
- Raise the trip current and document why. A 10 mA setting is appropriate for a small appliance, not for a 1 km cable.
- Use a DC hipot. A capacitor charges once and then draws nothing, so the leakage reading shows the insulation, not the capacitance.
Whichever you choose, record the ramp time and trip setting with the result. A hipot result without those two numbers cannot be reproduced.
Standards referenced: IEC 60243-1 for dielectric strength of insulating materials, IEC 61010-1 clause 6.8 for the test voltage and duration.
Reading the trip current correctly
The trip current on a hipot tester is the total current, which is the sum of resistive leakage and capacitive charging. During a ramp both are present; at the end of the ramp only the resistive part remains.
This is why a test that trips during the ramp but not at the hold is usually a capacitance story, and a test that trips only at the hold is usually a real insulation problem.
Setting the current limit for a cable
For a 1 km multicore cable with 100 nF of capacitance ramped at 1,500 V/s, charging current alone is 150 µA. Add a resistive leakage allowance of 1 mA at test voltage and the limit should be set at 2 mA, not 5 mA.
For a small appliance, the capacitance is negligible and 5 mA is a reasonable limit. The limit follows the load, not the tester.
What a DC test changes
A DC hipot charges the capacitance once and then draws only leakage. The charging current appears as a brief surge at the start, which the tester can ignore with a delay, and the steady reading is the insulation.
The trade-off is that DC stresses the insulation differently from AC. For cables and motors DC is usually acceptable; for equipment that will see AC in service, the AC test is the one that reflects reality.