The tip of a test probe is the part that does the measuring. It is also the part that takes the most wear, and the part most often inspected by eye. A tenth of a millimetre at the tip is enough to change a result.
Three kinds of tip wear
Shortening. The tip erodes or is ground back. The probe becomes effectively shorter, and a probe that should reach a barrier no longer does.
Rounding. A sharp tip becomes a dome. The rounded tip cannot enter a slot the original could, so a slot that should be accessible reads as protected.
Flattening. A dropped probe lands on its tip and acquires a flat. The flat presents a wider contact face and changes the entry geometry.
| Wear | What changes | Direction of error |
|---|---|---|
| shortening | reach | under-reports access |
| rounding | entry into narrow slots | under-reports access |
| flattening | contact area and angle | either, depending on orientation |
Which probes are affected most
Small-diameter probes: a 0.1 mm change on a 0.5 mm pin is 20 percent. The smaller the tip, the more a given wear matters.
Probes with a defined tip radius, like the cone and the thumb, where the radius is part of the definition. A rounded-off cone no longer represents the shape it models.
Probes used on abrasive surfaces — castings, rough mouldings — wear faster than probes used on smooth parts, because the contact is a point load on a hard edge.
How to measure
A toolmaker's microscope or a profile projector is the practical instrument. Measure the working length from the reference face to the tip, and the tip radius or flat diameter.
For routine checks, a go/no-go depth gauge made from a known-good probe is faster: if the worn probe sits lower in the gauge than the reference, it has shortened.
What is acceptable
No standard gives a wear limit for probe tips. The practical rule is derived from the tolerance being checked: if the tip wear is a significant fraction of the smallest feature the probe must enter, the probe is no longer fit.
For a 1 mm probe, 0.05 mm of shortening is 5 percent — usually acceptable. For a 0.5 mm pin, the same 0.05 mm is 10 percent, and the margin against the smallest slot it must pass is likely gone.
Recording
Working length and tip condition at each inspection, in numbers. "Tip worn" is not a record. "Working length 79.6 mm, tip radius 0.4 mm, limit 79.8 / 0.3" tells the next person whether to keep using it.